Skip to main content

Table 3 Diagnostic accuracy of 0.8% Immulab, Bio-rad, Grifols and QuidelOrtho RRBCs. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy of Immulab RRBCs compared to Bio-rad (test script 1), Grifols (test script 2) and QuidelOrtho (test script 3) 0.8% RRBCs with 95% confidence intervals (95% CI)

From: Evaluation of the analytical performance of four different manufacturer’s reagent red blood cells in antibody detection and identification

 

Test Script 1:

Immulab and Bio-Rad

Test Script 2:

Immulab and Grifols

Test Script 3:

Immulab and Ortho

 

Immulab 0.8% Abtectcell

III

Bio-Rad 0.8% ID-DiaCell I-II-III

Immulab 0.8% Abtectcell III

Grifols 0.8% Perfect Screen 3

Immulab 0.8% Abtectcell

III

QuidelOrtho 0.8% Surgiscreen

True-positive

72

69

80

66

69

66

False-positive

0

0

0

0

0

0

True-negative

15

15

8

8

24

24

False-negative

0

3

0

14

0

3

Sensitivity (%) (95% CI)

100

(95.01–100)

95.83

(88.30–99.13)

100

(95.49–100)

82.50

(72.38–90.09)

100

(94.79–100)

95.65

(87.82–99.09)

Specificity (%) (95% CI)

100

(78.20–100)

100

(78.20–100)

100

(63.06–100)

100

(63.06–100)

100

(85.75–100)

100

(85.75–100)

PPV (%)

(95% CI)

100

100

100

100

100

100

NPV (%)

(95% CI)

100

(78.20–100)

83.33

(58.58–96.42)

100

(63.06–100)

36.36

(17.20–59.34)

100

(85.75–100)

88.89

(70.84–97.65)

Accuracy (%)

(95% CI)

100

(95.85–99.97)

96.55

(90.25–99.28)

100.00

(95.89–100)

84.09

(74.75–91.02)

100

(96.11–100)

96.77

(90.86–99.33)