Skip to main content

Table 2 Samples with discrepant antibody identification results. Only the antibody specificities of samples which contained adequate volume to undergo complete antibody identification are shown

From: Evaluation of the analytical performance of four different manufacturer’s reagent red blood cells in antibody detection and identification

 

Test Script 1

(Bio-Rad CAT)

Test Script 2

(Grifols CAT)

Test Script 3

(Ortho CAT)

 

Immulab RRBCs

Bio-Rad RRBCs

Immulab

RRBCs

Grifols RRBCs

Immulab RRBCs

QuidelOrtho RRBCs

1

Anti-D

ND

    

2

Anti-E

ND

    

3

ND

Anti-Cw

    

4

Anti-M

ND

Anti-M

ND

Anti-M

ND

5

  

Anti-D

ND

  

6

  

Anti-D

ND

  

7

  

Anti-D

ND

  

8

  

Anti-D

ND

  

9

  

Anti-D

ND

  

10

  

Anti-E

ND

  

11

  

Anti-e

ND

  

12

  

Anti-Fya

ND

  

13

  

Anti-Jka

ND

  

14

  

Anti-Jka

ND

Anti-Jka

ND

15

  

Anti-M

ND

  

16

  

Anti-D + Anti-C

Anti-D

  

17

  

Anti-E + Anti-Kpa

Anti-Kpa

  

18

    

Anti-E

ND

  1. ND = no antibody detected